Beamline layout

Figure 3. URANOS beamline layout

Rysunek 4. Model linii URANOS

Figure 4. CAD model of the URANOS beamline.

The first element of the beamline that generates electromagnetic radiation is a quasiperiodic elliptically polarizing undulator that produces any polarized beam of radiation in the ultraviolet (UV)  and soft X-ray range. Then there are slits (horizontal and vertical) and a toroidal mirror M1 (16 m from the source), which collimates the beam vertically and focuses horizontally in the plane of the exit slit (26 m). And it leads the photon beam to the monochromator chamber (18.7 m) onto a mirror and diffraction grating system in a plane/grazing geometry (PGM - Plane Grating Monochromator) or a mirror and grating system in normal geometry (NIM - Normal Incidence Monochromator). The PGM monochromator works in the range of 12-600 eV, and the NIM in the range of 8-30 eV. Behind the monochromator chamber there is a cylindrical mirror (20 m) focusing the light in the vertical direction on the exit slit. Behind the exit slit is a toroidal mirror (27.3 m) that mirrors the image of the exit slit 1:1 on the sample surface in the analysis chamber at the end station.

Undulator

 

Rysunek 5. Undulator Apple II - źródło promieniowania linii URANOS.

Figure 5. Apple II undulator - source of radiation for the URANOS beamline.

The source of synchrotron radiation for the URANOS beamline is an elliptically polarizing Apple II type undulator with a quasiperiodic magnetic field geometry manufactured by Kyma SrL. The use of such a magnetic field geometry enables a significant improvement in the spectral purity of the UV beam (weakening of harmonic components). At the end of the beamline, the contribution of higher harmonics in the spectrum of excitation radiation falling on the sample is less than 1%. The elimination of harmonic components in the spectrum of excitation radiation is particularly important for the ARPES technique because the aim of this technique is to map the band structure of a solid. The intensity of the experimentally measured bands depends on the probabilities of the optical transitions in the solid, and there are sometimes small. Moreover, the number of observed bands, especially in materials with large, polyatomic lattice cells, can be very significant. In this situation, both the increase in the background level and the presence of band replicas caused by the presence of harmonic components in the spectrum of excitation radiation may prevent the correct interpretation of the measurement results.

The Apple II undulator can generate any polarized electromagnetic radiation; the following polarizations are available: linear horizontal, linear vertical, all linear skew, circular right and left, and any elliptical polarization.

Exemplary measurements of dichroic effects in ARPES, caused by the dependence between the matrix elements of optical transitions and the direction of electromagnetic field vibrations.

Widma ARPES ukazujące efekt dichroizmu kołowego na próbce Bi2Te3 dla różnych (odpowiednio kołowej lewoskrętnej i kołowej prawoskrętnej oraz różnica sygnałów) polaryzacji promieniowania o energii hv = 55eV.

Figure 6. ARPES spectra showing the effect of circular dichroism on a Bi2Te3 sample for different polarizations (left, right and signal difference, respectively) polarizations for energy excitation hv = 55eV.

Insert device parameters:

Parameter Value
Insert device type EPU Apple II, quasiperiodic, Kyma S.p.a.
Range of opening motion 20 (18) – 200 mm
Magnets 
Saturation field 
Magnetic field amplitude 
Number of periods 
Period length (λ0) 
Total length

NdFeB permanent magnets; 
1.24 T;
0.6 T, (0.8 T for 18 mm gap);

 

21;
120 mm;
2660 mm;

Mods Parallel and antiparallel
Polarization All, full control of light polarization: linear, horizontal, vertical, skew, circular right and left, any elliptical.
Radiation power at full closure 600 W (for current 500 mA)
Parametr K = e B λ μ 2 π m e c Max. 6.5 (for 18 mm gap)

 

Monochromator

 

Rysunek 7. Monochromator na linii URANOS

Figure 7. Monochromator chamber at URANOS beamline.

The monochromator chamber contains two optical systems; the first works in the plenary geometry, the grazing angle of incidence, PGM (plane grating monochromator), the second in the geometry of normal incidence, NIM (normal incidence monochromator). The PGM monochromator works with a grating with a line density of 600/mm, and the NIM with a grating with a line density of 2000/mm. These are the so-called laminar gratings, that is, those having a rectangular profile of grooves. The use of two different reflection geometries allows optimizing the parameters of the photon beam depending on the requirements: it can be, for example, maximization of the photon flux, maximum spectral purity of the beam, or minimal polarization deformation. The PGM monochromator works in the range of 14-600 eV, and the NIM works in the range of 8-30 eV.

PGM monochromator geometry. The figure shows two positions of the grid and the mirror relative to each other for the allowed angles of rotation.

Figure 8. PGM monochromator geometry. The figure shows two positions of the grid and the mirror relative to each other for the allowed angles of rotation.

Geometry of the NIM monochromator with a static mirror. Two extreme positions of the monochromator grating are shown.

Figure 9. Geometry of the NIM monochromator with a static mirror. Two extreme positions of the monochromator grating are shown.

The beam coming out of the monochromator must be directed towards the stationary exit slit (in the horizontal direction), i.e. regardless of the selected energy, the beam must always be at the same height relative to the exit slit.

The wavelength of the radiation coming out by the PGM and NIM monochromator can be determined by the formula:

λ = h c E = ( s i n β n + s i n α n ) N k

Where αn, βn are the angles of incidence and diffraction on the grating, respectively, measured with respect to the surface normal, N is the density of lines, grooves on the diffraction grating (for PGM 600 mm-1), and k is the order of diffraction.

For the grating working in the PGM mode, due to the variable angle of incidence (depending on the angle of the plane mirror), we also define the so-called constant cff (fix-focus constant):

c f f = c o s β n c o s α n

The cff parameter has practical use. If very high resolving power is required, this parameter should be increased (maximize, α and β have a limited range), while if we minimize harmonic components, then we should strive for the minimum possible cff values, as this will allow for more efficient suppression of harmonic components in the monochromator.

Energy resolution (RP), depending on the size of the exit slit Δs_ex , exit angle β_n, diffraction order, etc. is defined as:

R P = E Δ E = N k q λ Δ S e x c o s β n

where q defines the distance between the focusing mirror and the exit slit and is 6 m.