The method is based on the analysis of the distribution of the binding energies of the photoelectrons, emitted by the studied material upon excitation with ultraviolet radiation (ultraviolet photoelectron spectroscopy: UPS), or with X-ray radiation (X-ray photoelectron spectroscopy: XPS). If the distribution of the emission angles of the photoelectrons is also measured, the technique is called angle-resolved (angle-resolved photoelectron spectroscopy: ARPES)
XPS method, based on the excitation of core-level electrons, enables studying the chemical composition of materials. Because it is semi-quantitative, it allows to determine the stoichiometry of complex compounds. Chemical shifts of the emission maxima can be used to study the chemical environment of different atoms in the samples (type of chemical bonds, oxidation level). UPS method, based on the excitation of valence electrons, especially in the angle-resolved variant, allows for studying the electronic structure of various materials near the Fermi energy.
The method is surface-sensitive and the signal is collected from the layer with the depth of up to a few nanometers. Surface sensitivity can be increased, by tilting the sample, or by varying incoming photon energy. Under optimal circumstances, it is possible to obtain signal from a single atomic layer of the studied material.
PS method can be used in the following industries:
PS method is available on the following beamlines: URANOS and PHELIX . The basic parameters of the experimental stations and restrictions with regard to sample type are shown in table.
|
End station |
Type of samples |
Photon energy range |
|---|---|---|
|
UHV-compatible solids only |
8 – 600 eV |
|
|
UHV-compatible solids only |
30 – 1500 eV |